multi EA 5100
Operation
39
}
The basic module boots up. The LED on the front lights up in green after approx
30 s.
}
Start the multiWin program. Log in with user name and password.
}
Click the [Initialize analyzer] button.
ü
Initialization takes place after successful login.
NOTICE
Observe the run-in period
In the Status analyzer window, components that are not yet ready for operation are dis-
played in red. The heating time for 1050 °C is approx. 30 min. Measurement is not pos-
sible during the run-in period.
¡
If the analyzer is still not ready after approx. 30 min, perform a fault search (→
"Troubleshooting"
6.4
Switching off the basic module and the modules
NOTICE
Risk of overheating
If the basic module is switched off too early, the electronics can overheat and become
damaged due to lack of cooling.
¡
Only switch off the basic module after a cooling time of 1 h.
Switch off the basic modules and its modules as follows:
}
Exit the multiWin program.
}
Switch off the connected modules via their respective power switches.
}
Only switch off the basic module after a cooling time of 1 h.
If the system is switched off too early, the electronics inside the device can overheat
due to lack of cooling.
}
If a Cl module 5100 is connected: Before closing the gas supply, remove the sulfuric
acid container from the module and completely drain the sulfuric acid.
}
If a S module 5100 coulometric is connected: Before closing the gas supply, remove
the gas transfer line from the gas inlet tube of the measuring cell.
}
After switching off the modules, close the gas supply.
}
Exit the Windows operating system and switch off the PC.
ü
Shutdown of the basic modules and its modules is complete.