multi EA 5100
Specifications
177
Stirring mechanism of the measuring
cell
Integrated magnetic stirrer
(with fixed speed)
Operating range of the wide-range
coulometer
3
measuring cells
¡
"high sensitive"
¡
"sensitive"
¡
"high concentration"
The "high sensitive" measuring
cell
Measurement mode
Potentiometry
Measuring range (Cl absolute)
0.01 to 10 µg
Generator current
100 µA
Electrolyte volume
65 ml
The "sensitive" measuring cell
Measurement mode
Bi-amperometry
Measuring range (Cl absolute)
1 to 100 µg
Generator current
1 mA
Electrolyte volume
15 to 20 ml
The "high concentration" mea-
suring cell
Measurement mode
Bi-amperometry
Measuring range (Cl absolute)
10 to 1000 µg
Generator current
10 mA
Electrolyte volume
120 ml
Electrical variables
Power supply
110 to 240 V +10/-5 %
Overvoltage category
II
Frequency
50/60 Hz
Module fuse protection
Number of device fuses
T 2.0 A H
2
Typical average power consumption
50 VA
Interface to the basic module
RS 232
Only use original fuses from Analytik Jena!
General characteristics
Dimensions (W x H x D)
300 x 470 x 530 mm
Mass
12 kg
17.4 Technical data for the S module 5100 (basic, MPO)
Procedural data
Analytical parameters
Total sulfur TS
Detection principle
UV fluorescence
Measuring range (S in sample)
0.01 to 10000 mg/l S
Measuring range (S absolute)
0 to 100 µg S