multi EA 5100
Troubleshooting
109
¡
Check the cell temperature settings for the
method and adjust as necessary
Device not ready for measurement – Cl-POT
Display (green)
Description
End point routine
Automatic end point routine running:
¡
Wait for the end point routine to finish
Drift determination
Drift determination immediately after titration or end
point routine:
¡
Wait until drift determination is complete (ap-
prox. 1 minute)
Device not ready for measurement - CI-AMP
Display (red)
Description
Inactive
No cell detected:
¡
Insert cell
¡
Basic initialization
Device not ready for measurement - CI-AMP
Display (green)
Description
Stand-by titration
Interval titration running:
¡
Measurement start possible
End point routine
End point routine running:
¡
Wait for the end point routine to finish
Device not ready for measurement – C-NDIR
Display (red)
Description
Analog values warning
Analog values outside range:
¡
Access the values via the System |Component
test |NDIR menu item.
Running-in time
Detector not yet ready for operation:
¡
Wait for the run-in period after activation to com-
plete (approx. 30 min)
Device not ready for measurement – S-UVFD or N-CLD
Display (red)
Description
Preheating phase
Detector not yet ready for operation:
¡
Wait for the run-in period after activation to com-
plete (approx. 30 min)
Vacuum/pressure error
Pressure in the detector outside the permitted range:
¡
See device faults of the N detector
13.5 Device faults on the basic module
Error
Possible cause
Remedy
Furnace does not heat
Thermocouple connector
not connected
Connect the connector
(→ "Removing and in-
stalling the combustion
furnace"
Temperature set incor-
rectly in the software
Check temperature con-
figuration in the method