multi EA 5100
Maintenance and care
153
NOTICE
Contamination reduces the effectiveness of the UV lamp.
¡
Do not touch the glass of the new lamp. In particular, protect the radiation emission
point made of quartz glass.
¡
If the glass is touched, wipe it with a lint-free cloth and pure alcohol.
The following states indicate a defective UV lamp:
¡
The run-in period of the detection module takes is not complete after 30 min. The
LED on the front flashes continuously.
¡
The measurement sensitivity is too low or the detection limit can no longer be
reached.
¡
Check the state of the UV lamp in the System |Component test menu in the detec-
tion module tab. If a defect is indicated or the lamp is worn out, replace the lamp.
}
Switch off the detection module via the power switch.
}
Remove the left side panel. Remove the 4 screws to do this. Remove the
protective grounding cable and remove the side panel.
ü
The UV lamp is located to the left in the module (see arrow).
}
Unscrew both fastening screws with a Phillips screwdriver.
}
Pull the connection plug on top out of the socket.
}
Carefully remove the lamp from its holder.
}
Insert a new lamp in the holder.
NOTICE! Only touch the new lamp by its base or its cable. Do not touch
the glass. Do not scratch the lamp.