Troubleshooting
multi EA 5100
100
Error
code
Error message/cause
Remedy
100
101
C sensor: No connection
C sensor: CRC error
Communication interrupted after sensor detection
during initialization
Acknowledge message
Initialize
104
C sensor: Analog values outside range
The analog values of the detector are outside the
operating range
Check the quality of the carrier gas
Initialize
Check the analog values in the component test
(System |Component test |NDIR menu item)
110
120
130
N sensor: No connection
S sensor: No connection
Cl sensor: No connection
Communication interrupted after sensor detection
during initialization
Acknowledge message
Initialize
131
Cl sensor: Incorrect command structure
Communication to the chlorine module inter-
rupted
Acknowledge message
Switch the chlorine module on/off
Initialize
132
Cl sensor Indicator error
Indicator value out of range after titration start
(measurement cannot start)
Acknowledge message
Initialize
Run an end point routine
Check the status of the titration cell (System |
Component test |Chlorine menu item)
133
Cl sensor: Incorrect cell
No initialization after cell change
Initialize
134
Cl sensor: Incorrect status
Communication error
Acknowledge message
Initialize
Check the status of the titration cell ( System |
Component test | Chlorine menu item)
135
Cl sensor: Incorrect version
Transmission fault
Acknowledge message
Initialize
Check the status of the titration cell ( System |
Component test | Chlorine menu item)
200
Gas box: No connection
Communication error
Acknowledge message
Initialize