multi EA 5100
Chlorine analysis with the Cl module 5100
57
1
2
3
4
5
Figure 37Measuring cell connections
1 Combined electrode connection
2 Measuring cells "sensitive" and
"high concentration" connection
3 Platinum electrode connection
4 Silver electrode connection
5 Sensor and reference electrode con-
nection (via pre-amplifier)
Measuring gas transfer
A heated gas transfer line is used to transfer the measuring gas. It connects the combus-
tion system of the basic module with the sulfuric acid container in the detection module.
The hose of the gas transfer line is connected to the sulfuric acid container connection
with a banjo bolt with a conical nipple. The other end of the gas transfer line is con-
nected to the auto-protection valve assembly in the basic device.
1
2
3
4
5
6
7
3
5
4
Figure 38Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment
2 Sulfuric acid container
3 Hose for measuring gas infeed
4 Connector
5 Banjo bolt
6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector
The measuring gas is routed into the measuring cell via the safety attachment of the
sulfuric acid container.
¡
The measuring gas hose is connected directly to the combined electrode with the
"sensitive" and "high concentration" measuring cells.