multi EA 5100
Troubleshooting
123
Error
Possible cause
Remedy
Low results
¡
Catalyst depleted
¡
Exchange the catalyst
¡
System is leaking
¡
Check the sluice for leaks
¡
Replace septum
¡
Incorrect injection vol-
ume
¡
For manual sampling:
introduce the sample
volume set in the
method
¡
Phosphoric acid in the
TIC reactor depleted
¡
Regenerate the TIC reac-
tor, see (→ "Regenerat-
ing the TIC reactor"
¡
Septum faulty
¡
Replace septum