multi EA 5100
Troubleshooting
107
The following status displays are possible:
Device not ready for measurement
Display
Description
No sampling module dis-
played (display empty)
Device not ready for measurement, no sampling mod-
ule identified:
¡
Connect and activate at least one sampling mod-
ule
¡
Basic initialization
Auto-injector - syringe: ?
No auto-injector syringe detected
¡
Initialize
¡
Insert auto-injector syringe and register it in mul-
tiWin ("Extras | New auto-injector syringe" menu
item)
13.4.4 Detectors
Example
Cl-POT OK
3050
Meaning
Detector status
Current sensor value
All detector modules detected during initialization are displayed here. The following dis-
plays are possible:
Display
Description
C-NDIR
C module 5100 or TOC module 5100 with NDIR detector
N-CLD
N module 5100 with chemiluminescence detector
S-UVFD
S module 5100 basic and S module 5100 MPO with UV
flourescence detector
S-Coul
S module 5100 coulometric with microcoulometer
Cl-POT
Cl module 5100 with "high sensitive" measuring cell
Cl-AMP smallCell
Cl module 5100 with "sensitive" measuring cell
Cl-AMP largeCell
Cl module 5100 with "high concentration" measuring cell
The respective status of the detector module is indicated by the color:
Display
Description
Black
Detector is active, status is queried and displayed (see example
above)
Gray
Detector is inactive, status is not displayed
Green
Detector OK, device ready for measurement (OK)
Or
Detector busy, measurement can only be started once the routine is
complete (detector-specific)
Red
Error, see overview below