3.3.2 WATER OUT Wastewater connection
• Connect the red wastewater hose to the orange wastewater connector that is marked on the rear panel
of the WATER OUT device. The tubing should be inserted into the connector, to the depth of the black
line (placed on the tubing).
• Wastewater should be discharged directly to the water and sewage system or to a special wastewater
tank. If a tank is used, the other end of the red tube should be placed in the wastewater tank. To
secure and immobilize the water drain hose from the machine, use the plug provided with the hose
and insert the plug into the opening of the wastewater tank. The unsecured tubing may jump out of the
tank and flood the room.
• Make sure that the red tube is never immersed in water, otherwise the water will not drain properly,
causing sterilization errors.
• The wastewater tank or drainage outlet must always be located below the unit.
• When using wastewater tanks, we recommend that you use tanks with the same capacity as the feed
water tank. Their simultaneous replacement will protect against the possibility of wastewater overflow.
• The water level in the tank should be checked on a regular basis, depending on the frequency of the
processes.
3.3.1 Connecting the WATER IN feed water
• Connect the blue connection hose to the blue feedwater connection, which is marked on the rear panel
of the WATER IN device. The tubing should be inserted into the connector, to the depth of the black
line (placed on the tubing).
• The other end of the blue tube should be immersed in the tank with the feed water. The device is
equipped with a water suction pump, so it is not necessary to place the water tank above or on the
same level as the device.
• To secure and immobilize the water supply hose to the machine, use the plug supplied with the hose
and insert the plug into the opening of the water supply tank. An unsecured tubing may jump out of the
water and cause sterilization errors.
• To prevent the hose from coiling in the water tank, install the included weight with a rubber ring at 2
cm from the end of the hose (fig.)
• The minimum water load in the tank is 300 ml.
• Remember and check that the blue tube is always immersed in the water.
• Check the water level in the tank on a regular basis, depending on the frequency of the processes
3.3. Water connection with feed and wastewater tank
To connect the Enbio autoclave, you need 2 water tanks (feed and waste) of identical capacity, or a feed
water tank and drainage access to the sewage system.
Correct positioning of the hoses in
the feed and wastewater tanks.
Correct connection of
the hoses to the feed and
wastewater connections
Correct installation of the
weight with the securing ring
20 m
m
14