Replacement of consumables
The sterilizer is equipped with a high performance sterilization system. A message on the
screen informs the user when each component should be replaced. If the sterilizer is in
regular operation, alternating replacement screens will appear after the welcome screen
has been pressed. The replacement screens are described in detail in section "9. Warning
Messages and Error Codes".
Cleaning the water container
In order to ensure the correct parameters of the water supplying the device, it is
recommended to check the water tank at least quarterly. If contamination is found, the tank
should be emptied, cleaned and refilled with new water.
Replacing the Enbio Magic Filter
The filter should be replaced with a new one once every 6 months or sooner when the resin has
completely discolored to an amber color. Failure to do so may disrupt the sterilization process and
will void the sterilizer warranty. Dispose of the used filter according to local guidelines. The filter and
all its parts are not recyclable.
In order to ensure an efficient sterilization process and the correct functioning of the
device, it is recommended to replace consumables in a timely manner.
New filter
Filter half
used up
Need to replace
6.1 Consumables
The table below contains items subject to periodic replacement and items subject to natural wear. Spare
parts must be ordered directly from the manufacturer. The use of other spare parts voids the warranty and
does not guarantee the correct functioning of the device.
Name
Part no
HEPA filter Enbio S / PRO
1-8-27720A5
Enbio S front kit with gasket
1-8-14672D2
Enbio S tray
1-8-14738B2
Enbio PRO front kit with gasket
1-8-12433A45
Tray set 3 pieces Enbio PRO
1-8-1097954A1
Wastewater set (plug + red hose 1.5 m)
1-8-42988A1
Feed water set (plug + blue hose 1.5 m)
1-8-42992A1
Enbio Magic Filter
1-8-1116687A5
Cleaning the gasket
It is recommended to clean the gasket each time after 100 processes. Use warm, clean water
and microfiber to clean the gasket (microfiber with silver particles is allowed). Blunt and sharp
cleaning tools are not allowed. Cleaning with chemicals is not allowed. Carry out cleaning
on a cooled down device, after opening the drawer. Be careful not to bend the drawer. After
cleaning, leave the device open until the gasket is dry. During this time, protect the device
against damage.