The counter can only be reset by Enbio or an authorized external service
by selecting the RESET button on the counter screen and entering a
unique service code.
Displaying the counter values in yellow or red does not block the operation of the
unit. However, exceeding the required replacement time may have a significant
impact on the operation of the unit and the sterility of the load. For replacement of
individual components, please contact the manufacturer or supplier.
No.
Name
Replacement
frequency
(cycles)
Yellow
Early replacement call
(cycles)
Red
Latest replacement after
(cycles)
1
HEPA filter
1000
980
1000
5.5 Restart
The restart of the process is forced when the user stops the process by pressing the STOP field, in the
case of a power outage or an error during the process, for example lack of supply water.
If the STOP field is selected, the following messages are displayed alternately:
• the user has stopped the process
• equalizing the pressure in the working chamber
• process is incorrect, which means that the load is not sterile.
When the pressure in the working chamber is equalized, the following messages appear alternately on the
display. Now you can open the device freely. After opening the chamber, the screen appears.
By selecting the field, we can return to the welcome screen. In the event of an error, we must additionally
enter the 4-digit security code 0000. Entering this code is tantamount to the operator's declaration that he
is aware that the sterilization process has not been carried out properly and that the batch is non-sterile.