5.3 Test programs
If a USB flash drive has not been inserted into the
unit, the USB symbol is not displayed in the lower right
corner of the screen and a message indicating no USB
flash drive is displayed. The test program data will not
be saved. We can continue without writing data to the
USB flash drive by pressing the YES box or abort by
selecting the NO box to place the USB flash drive in
the port and start the program from the beginning.
By pressing the Test field you go to the test programs
menu. From this level, we can choose between the
Vacuum test and Helix / B&D test programs. We select
the appropriate program by pressing the required field
on the display.
When the device working chamber is closed, the word
DOOR OPEN changes to START and by pressing this
field the selected test program is started.
Enbio S / Enbio PRO
Bowie & Dick / Helix
Vacuum test
Process temperature
134°C
–
Number of pre-vacuums
3
1
Sterilization time
3.5 min
–
Drying time
3 min
–
Total process time
15 min
16 min
Vacuum test
The vacuum test should only be performed on a cold device before starting work. During the test, the
device verifies:
• vacuum pump capacity.
• tightness of the pneumatic system.
During the vacuum test the sterilizer chamber must be completely dry and cold. If not, the vacuum
test may not be reliable even if the sterilizer is fully operational. When the test is complete a
message indicating the result will appear on the display. If the result is negative check, clean or
replace the seal, clean the front edge of the chamber and repeat the test. If the Vacuum test fails
again, contact the supplier or manufacturer.
After selecting a vacuum test program and starting
it with the button, the vacuum test program run
screen is displayed, with Information about loss of
pressure in the working chamber and the duration
of the test.
[Vacuum test PASSED]
When the test program ran successfully.
[Vacuum test FAILED]
When the test program did not run successfully.
After pressing the CONTINUE field, the welcome
screen is displayed.
The devices are equipped with special test programs to
verify the correct operation of the sterilizer.
26
27