ENBIO Enbio Pro Скачать руководство пользователя страница 27

Messages

"Aborted by user"

Process interrupted by the 
user. Non-sterile cartridge if 
interrupted during or before 
the sterilization process.

This message appears when the user 
terminates the process. This does not 
mean that there is a malfunction. Start a 
new process.

"Vacuum test failed"

Vacuum test error

Contact with the service

"No USB memory"

No USB memory stick

Check the USB port and mount the 
memory. Contact with the service.

"Equalizing pressure"   

Pressure when stationary. 
Equalization of pressure to 
atmospheric.

•  The message occurs in certain cases 

as a result of natural processes. 

•  If the message appears frequently, 

contact the service center.

"Overpressure during 
standby"

Hypertension in standby 
mode

The reason for this error is that the 
hot sterilizer is left with the chamber 
closed (e.g. overnight). As the sterilizer 
cools down a vacuum is created in the 
chamber which causes a startup error. 
Wait until the device has equalized the 
pressure automatically - the message will 
disappear automatically

10 "Chamber heating error"

Chamber heating error

• 

Mains voltage too low - consult an 

electrician for the location where the 

autoclave is to be installed

• 

Contact with the service

11

"Steam generator heating 

error"

Steam generator error

• 

Repeat the process

• 

Contact with the service

12 "Prevacuum fail / check 

outlet"

Vacuum pump / drain fault

• 

Check the level of used water in the bottle 

with the red cap and pour out the excess

• 

Check that the red hose is not submerged 

or kinked

• 

Check that the autoclave setting provides 

free airflow for cooling the unit 

• 

The red hose must point downwards 

along its entire length, no section may 

point upwards

• 

Clean the chamber door seal

• 

Contact with the service

13 "Power failure"

Momentary voltage loss during 

operation

• 

Restart the device and make sure it is 

properly plugged into the outlet

• 

Consult an electrician competent for the 

site where the autoclave is installed to 

check the installation

14 "Pressure during standby" Overpressure while waiting  

• 

Restart your device

• 

Contact with the service 

15 "Locking door error"   

Door lock error

Contact with the service

16 "Unlocking door error"

Door unlock error

• 

Turn off the autoclave and turn it on 

again, start the process and stop it 

after a few seconds. There must be no 

overpressure in the chamber, i.e. the 

following information must be displayed: 

"READY / Chamber is safe to open"

• 

Contact with the service

17 "Valve V3 / HEPA filter 

error"

V3 valve / HEPA filter error

• 

Replace the HEPA filter

• 

Contact with the service

18 "Pressure sensor error"

Pressure sensor error 

Contact with the service

19 "USB disc error / Change 

disc"

Writing error on pendrive - 

damage to medium

Copy the contents from your current flash 

drive - buy and install a new one

20 Min. Chamber temperature Chamber temperature too low 

during the process

Contact with the service

21 Chamber temperature 

sensor failure

Chamber temperature sensor 

failure

Contact with the service

22 Steam gen. Temp. Sensor 

failure

Steam generator temperature 

sensor failure

Contact with the service

23 Process temp. Sensor 

failure

Process temperature sensor 

failure

Contact with the service

24 Autoclave has too low 

temperature

Autoclave temperature too low / 

temperature sensor error

• 

Leave the device switched off for 3 hours 

at room temperature

• 

Contact with the service

31 "Internal flash error"   

Internal memory error

Contact with the service

50

51

Содержание Enbio Pro

Страница 1: ...2022 04 27 Enbio S Enbio PRO User Manual The latest version of the manual is available at www enbio com EN...

Страница 2: ...ging packs on sterilizer tray sterilizing packaged loads 4 3 Arranging the packages on the sterilizer tray sterilization of loads without packaging 5 Operating the device 5 1 First run 5 2 Program sel...

Страница 3: ...are suitable for sterilizing loads for which steam sterilization is specified The following medical devices can be sterilized in the ENBIO S and ENBIO PRO solid loads small porous items small porous...

Страница 4: ...multiple layers 134 C 4 minutes 3 minutes 500 grams 3 load of 100 grams 13 minutes 115 ml B 121 C solid loads small porous objects small porous loads full porous loads simple dimpled objects narrow c...

Страница 5: ...al process time for a load of specified weight Max water consumption Sterilization process class 134 C solid loads small porous objects small porous loads full porous loads simple dimpled objects narr...

Страница 6: ...Representative The correct and safe operation of the device is based on a thorough reading and observance of this document on the installation and use of the device in accordance with the description...

Страница 7: ...ure and humidity different from that in the place of installation wait at least 60 minutes after delivery If the device is moved from a cold to a warm place it may contain humidity which by adversely...

Страница 8: ...sterilization water must be less than 10 ppm or in the case of conductivity measurement it must be less than 15 S cm therefore tap water cannot be used as feed water for the device The table below sh...

Страница 9: ...nect the blue connection hose to the blue feedwater connection which is marked on the rear panel of the WATER IN device The tubing should be inserted into the connector to the depth of the black line...

Страница 10: ...of feedwater WATER IN with Enbio Magic Filter Connect the shorter part of the blue Enbio Magic Filter connection hose to the blue water supply connection which is marked on the rear panel of the WATER...

Страница 11: ...or other places where the unit could be flooded danger of short circuit The device should be installed in a well ventilated room far from heat sources and rooms where mixtures of gases or liquids or...

Страница 12: ...the packaging material must not be placed too loosely or be tightly stretched so as not to affect pressure variations during sterilization a label with information about the contents of the package th...

Страница 13: ...IO PRO device does not affect the biocompatibility of the materials All device components that come into direct contact with the sterilized load do not cause toxicity sensitization or irritation Proce...

Страница 14: ...rner of the screen During the program run the screen displays the temperature of the selected sterilization program or the current temperature in the process chamber in the lower left corner the curre...

Страница 15: ...tal process time 15 min 16 min Vacuum test The vacuum test should only be performed on a cold device before starting work During the test the device verifies vacuum pump capacity tightness of the pneu...

Страница 16: ...x B D test program can be stopped at any time by pressing the STOP field which is associated with incorrect termination of the test When the test program is complete alternating screens are displayed...

Страница 17: ...ly check that the tubing is not clogged and obstructed and repeat the test WARNING The test kit will be hot To correctly interpret the test refer to the instructions provided by the test manufacturer...

Страница 18: ...user can reset the values by pressing the RESET button The value for the service check can only be reset by an authorized service technician During the first process Vacuum Helix FAST 134 121 the uni...

Страница 19: ...ps the process by pressing the STOP field in the case of a power outage or an error during the process for example lack of supply water If the STOP field is selected the following messages are display...

Страница 20: ...tion of the device It is recommended to clean the inside of the process chamber once a week with a mild detergent without the addition of chlorine After cleaning the chamber should be wiped dry with a...

Страница 21: ...n process and the correct functioning of the device it is recommended to replace consumables in a timely manner New filter Filter half used up Need to replace 6 1 Consumables The table below contains...

Страница 22: ...g guidelines Name Replacement frequency HEPA bacteriological filter Every 1000 cycles or every 12 months Connection drain hose If damage is observed or once a year Plugs for water condensate container...

Страница 23: ...on can be downloaded from the manufacturer s website http enbio com service 8 1 Installing the software To install the software double click on the software installation file After this operation an i...

Страница 24: ...n window consists of three main areas Tree of all processes that have been synchronized with the flash drive have been sorted by execution dates Temperature and pressure graph with main autoclave and...

Страница 25: ...he program and its version Search The program allows you to search for processes by date range process number process type process outcome success fail PDF report The program allows you to generate a...

Страница 26: ...tle with the blue hose Correct the position of the blue hose so that the end is completely submerged in water Add a sinker to eliminate the problem in the future Check that the water supply hose blue...

Страница 27: ...rds along its entire length no section may point upwards Clean the chamber door seal Contact with the service 13 Power failure Momentary voltage loss during operation Restart the device and make sure...

Страница 28: ...ort enbio com WARNING The complaint process will be started when the service department receives a correctly completed complaint application When returning the device for service clean the chamber and...

Страница 29: ...ange from 5 C to 25 C Relative humidity 0 90 Storage temperature range 20 C to 60 C Relative humidity 0 90 Ambient pressure range 900 1100 hPa Rating plate located on the bottom of the device Enbio S...

Страница 30: ...June 1993 concerning medical devices as amended The device has been classified in Class IIb in accordance with rule 15 of Annex IX of the above mentioned Directive Conformity assessment has been carri...