Messages
"Aborted by user"
Process interrupted by the
user. Non-sterile cartridge if
interrupted during or before
the sterilization process.
This message appears when the user
terminates the process. This does not
mean that there is a malfunction. Start a
new process.
"Vacuum test failed"
Vacuum test error
Contact with the service
"No USB memory"
No USB memory stick
Check the USB port and mount the
memory. Contact with the service.
"Equalizing pressure"
Pressure when stationary.
Equalization of pressure to
atmospheric.
• The message occurs in certain cases
as a result of natural processes.
• If the message appears frequently,
contact the service center.
"Overpressure during
standby"
Hypertension in standby
mode
The reason for this error is that the
hot sterilizer is left with the chamber
closed (e.g. overnight). As the sterilizer
cools down a vacuum is created in the
chamber which causes a startup error.
Wait until the device has equalized the
pressure automatically - the message will
disappear automatically
10 "Chamber heating error"
Chamber heating error
•
Mains voltage too low - consult an
electrician for the location where the
autoclave is to be installed
•
Contact with the service
11
"Steam generator heating
error"
Steam generator error
•
Repeat the process
•
Contact with the service
12 "Prevacuum fail / check
outlet"
Vacuum pump / drain fault
•
Check the level of used water in the bottle
with the red cap and pour out the excess
•
Check that the red hose is not submerged
or kinked
•
Check that the autoclave setting provides
free airflow for cooling the unit
•
The red hose must point downwards
along its entire length, no section may
point upwards
•
Clean the chamber door seal
•
Contact with the service
13 "Power failure"
Momentary voltage loss during
operation
•
Restart the device and make sure it is
properly plugged into the outlet
•
Consult an electrician competent for the
site where the autoclave is installed to
check the installation
14 "Pressure during standby" Overpressure while waiting
•
Restart your device
•
Contact with the service
15 "Locking door error"
Door lock error
Contact with the service
16 "Unlocking door error"
Door unlock error
•
Turn off the autoclave and turn it on
again, start the process and stop it
after a few seconds. There must be no
overpressure in the chamber, i.e. the
following information must be displayed:
"READY / Chamber is safe to open"
•
Contact with the service
17 "Valve V3 / HEPA filter
error"
V3 valve / HEPA filter error
•
Replace the HEPA filter
•
Contact with the service
18 "Pressure sensor error"
Pressure sensor error
Contact with the service
19 "USB disc error / Change
disc"
Writing error on pendrive -
damage to medium
Copy the contents from your current flash
drive - buy and install a new one
20 Min. Chamber temperature Chamber temperature too low
during the process
Contact with the service
21 Chamber temperature
sensor failure
Chamber temperature sensor
failure
Contact with the service
22 Steam gen. Temp. Sensor
failure
Steam generator temperature
sensor failure
Contact with the service
23 Process temp. Sensor
failure
Process temperature sensor
failure
Contact with the service
24 Autoclave has too low
temperature
Autoclave temperature too low /
temperature sensor error
•
Leave the device switched off for 3 hours
at room temperature
•
Contact with the service
31 "Internal flash error"
Internal memory error
Contact with the service