1.5. Enbio S sterilization parameters
The table below, which presents the characteristics of the individual sterilization
programs in the Enbio S device, should be read in detail. It is the user's
responsibility to select the program appropriate to the type of load and not to
exceed the maximum weight of the sterilized load. Failure to comply with the
following rules may endanger the health of patients and the operation of the device.
Enbio S
Program name
Description of the load
Sterilization
temperature
Sterilization
time
Drying time
Max. load
weight
Number of
pre-vacuums
Total process
time (for a load of
specified weight)*
Max. water
consumption
Sterilization
process
class
134˚C
solid loads, small porous objects, small porous loads, full porous loads,
simple dimpled objects, narrow clearance objects, multiple packages that
may be unpackaged or packaged (single or multiple layers)
134˚C
4 minutes
3 minutes
500 grams 3
load of 100 grams
≈ 13 minutes
115 ml
B
121°C
solid loads, small porous objects, small porous loads, full porous loads,
simple dimpled objects, narrow clearance objects, multiple packages that
may be unpackaged or packaged (single or multiple layers)
121°C
15 minutes
5 minutes
500 grams 3
load of 100 grams
≈ 26 minutes
110 ml
B
134°C FAST **
solid, non-porous, simple dental instruments and appliances (e.g. scissors,
handles, pliers, chisels, probes, etc.) only unpackaged, non-textile
134˚C
3.5 minutes
no drying
500 grams 1
load of 100 grams
≈ 7 minutes
105 ml
S
* Ambient temperature can have an effect on increasing the total process
time.
* The total duration of the first process may be longer due to the need for the
appliance chamber to warm up.
** After sterilization at 134°C FAST the loads are wet. It is recommended to
leave the drawer ajar in order to evaporate the remaining water and to lower
the temperature of the instruments.
** After sterilization in the 134°C FAST program, the instruments are intended
for immediate use.