3.4.   Connection with water recycling (Enbio Magic Filter)

Enbio Magic Filter is a filter for recycling water used in the sterilizer. The ion-exchange resin contained 
in Enbio Magic Filter allows you to purify wastewater to the level of deionized water in accordance with 
the requirements of EN 13060 - that is, water which is recommended and approved for use with Enbio 
autoclaves. You only need 1 water tank to use Enbio autoclave with filter.

3.4.2  WATER OUT wastewater connection with Enbio Magic Filter

•  Connect the red wastewater hose to the orange wastewater connector that is marked on the rear panel 

of the WATER OUT device. The tubing should be inserted into the connector, to the depth of the black 
line (placed on the tubing).

•  Wastewater should be drained into a water tank. To secure and immobilize the water drain hose from 

the unit, use the plug supplied with Enbio Magic Filter and insert the plug into the opening of the water 
tank. The unsecured tubing may jump out of the tank and flood the room. 

•  Make sure that the red tube is never immersed in water, otherwise the water will not drain properly, 

causing sterilization errors. 

•  The water tank must always be located below the unit.

3.4.1   Connection of feedwater WATER IN with Enbio Magic Filter

•  Connect the shorter part of the blue Enbio Magic Filter connection hose to the blue water supply 

connection, which is marked on the rear panel of the WATER IN device. The tubing should be inserted 
into the connector, to the depth of the black line (placed on the tubing).

•  Submerge the longer part of the blue Enbio Magic Filter hose into the feed water tank. The unit is 

equipped with a water intake pump, so there is no need to place the water tank above or at the same 
level as the unit 

•  To ensure proper operation, Enbio Magic Filter must always be placed in a vertical position
•  To prevent the hose from coiling in the water tank, install the included weight with a rubber ring at 2 

cm from the end of the hose (fig.)

•  To secure and immobilize the water supply hose to the machine, use the plug supplied with the Enbio 

Magic Filter and insert the plug into the opening of the water tank. An unsecured tubing may jump out 
of the water and cause sterilization errors. 

•  The minimum water load in the tank is 300 ml. 
•  Remember and check that the blue tube is always immersed in the water.
•  Check the water level in the tank on a regular basis, depending on the frequency of the processes

Correct connection of 
the hoses to the feed and 
wastewater connections

Correct installation of the 
weight with the securing ring

Correct positioning of the 
hoses in the water tank

16

17

Содержание Enbio Pro

Страница 1: ...2022 04 27 Enbio S Enbio PRO User Manual The latest version of the manual is available at www enbio com EN...

Страница 2: ...ging packs on sterilizer tray sterilizing packaged loads 4 3 Arranging the packages on the sterilizer tray sterilization of loads without packaging 5 Operating the device 5 1 First run 5 2 Program sel...

Страница 3: ...are suitable for sterilizing loads for which steam sterilization is specified The following medical devices can be sterilized in the ENBIO S and ENBIO PRO solid loads small porous items small porous...

Страница 4: ...multiple layers 134 C 4 minutes 3 minutes 500 grams 3 load of 100 grams 13 minutes 115 ml B 121 C solid loads small porous objects small porous loads full porous loads simple dimpled objects narrow c...

Страница 5: ...al process time for a load of specified weight Max water consumption Sterilization process class 134 C solid loads small porous objects small porous loads full porous loads simple dimpled objects narr...

Страница 6: ...Representative The correct and safe operation of the device is based on a thorough reading and observance of this document on the installation and use of the device in accordance with the description...

Страница 7: ...ure and humidity different from that in the place of installation wait at least 60 minutes after delivery If the device is moved from a cold to a warm place it may contain humidity which by adversely...

Страница 8: ...sterilization water must be less than 10 ppm or in the case of conductivity measurement it must be less than 15 S cm therefore tap water cannot be used as feed water for the device The table below sh...

Страница 9: ...nect the blue connection hose to the blue feedwater connection which is marked on the rear panel of the WATER IN device The tubing should be inserted into the connector to the depth of the black line...

Страница 10: ...of feedwater WATER IN with Enbio Magic Filter Connect the shorter part of the blue Enbio Magic Filter connection hose to the blue water supply connection which is marked on the rear panel of the WATER...

Страница 11: ...or other places where the unit could be flooded danger of short circuit The device should be installed in a well ventilated room far from heat sources and rooms where mixtures of gases or liquids or...

Страница 12: ...the packaging material must not be placed too loosely or be tightly stretched so as not to affect pressure variations during sterilization a label with information about the contents of the package th...

Страница 13: ...IO PRO device does not affect the biocompatibility of the materials All device components that come into direct contact with the sterilized load do not cause toxicity sensitization or irritation Proce...

Страница 14: ...rner of the screen During the program run the screen displays the temperature of the selected sterilization program or the current temperature in the process chamber in the lower left corner the curre...

Страница 15: ...tal process time 15 min 16 min Vacuum test The vacuum test should only be performed on a cold device before starting work During the test the device verifies vacuum pump capacity tightness of the pneu...

Страница 16: ...x B D test program can be stopped at any time by pressing the STOP field which is associated with incorrect termination of the test When the test program is complete alternating screens are displayed...

Страница 17: ...ly check that the tubing is not clogged and obstructed and repeat the test WARNING The test kit will be hot To correctly interpret the test refer to the instructions provided by the test manufacturer...

Страница 18: ...user can reset the values by pressing the RESET button The value for the service check can only be reset by an authorized service technician During the first process Vacuum Helix FAST 134 121 the uni...

Страница 19: ...ps the process by pressing the STOP field in the case of a power outage or an error during the process for example lack of supply water If the STOP field is selected the following messages are display...

Страница 20: ...tion of the device It is recommended to clean the inside of the process chamber once a week with a mild detergent without the addition of chlorine After cleaning the chamber should be wiped dry with a...

Страница 21: ...n process and the correct functioning of the device it is recommended to replace consumables in a timely manner New filter Filter half used up Need to replace 6 1 Consumables The table below contains...

Страница 22: ...g guidelines Name Replacement frequency HEPA bacteriological filter Every 1000 cycles or every 12 months Connection drain hose If damage is observed or once a year Plugs for water condensate container...

Страница 23: ...on can be downloaded from the manufacturer s website http enbio com service 8 1 Installing the software To install the software double click on the software installation file After this operation an i...

Страница 24: ...n window consists of three main areas Tree of all processes that have been synchronized with the flash drive have been sorted by execution dates Temperature and pressure graph with main autoclave and...

Страница 25: ...he program and its version Search The program allows you to search for processes by date range process number process type process outcome success fail PDF report The program allows you to generate a...

Страница 26: ...tle with the blue hose Correct the position of the blue hose so that the end is completely submerged in water Add a sinker to eliminate the problem in the future Check that the water supply hose blue...

Страница 27: ...rds along its entire length no section may point upwards Clean the chamber door seal Contact with the service 13 Power failure Momentary voltage loss during operation Restart the device and make sure...

Страница 28: ...ort enbio com WARNING The complaint process will be started when the service department receives a correctly completed complaint application When returning the device for service clean the chamber and...

Страница 29: ...ange from 5 C to 25 C Relative humidity 0 90 Storage temperature range 20 C to 60 C Relative humidity 0 90 Ambient pressure range 900 1100 hPa Rating plate located on the bottom of the device Enbio S...

Страница 30: ...June 1993 concerning medical devices as amended The device has been classified in Class IIb in accordance with rule 15 of Annex IX of the above mentioned Directive Conformity assessment has been carri...

Отзывы: