When the chamber is closed by turning the
locking knob clockwise, the START symbol
appears on the display, which informs that the
chamber is properly closed.
Now we can select the program by pressing the appropriate
symbol of the temperature in which we want to sterilize 121 ° C,
134 ° C, 134 ° C FAST or 134 ° C PRION, which will highlight the
selected program. The selected program is started by pressing
the START symbol.
If the USB memory is not inserted in the device,
the USB disk symbol is not displayed in the lower
right corner of the screen.
A message appears that USB memory is missing.
The program data will not be saved. We can
continue working without saving data to the USB
memory stick by pressing the YES box, or abort
the work by selecting the NO box to place the
stick in the port and start the program steps from
the beginning.
It is recommended to use a USB drive with each
of the processes. The data stored on it will allow
you to print reports from selected processes.
If work continues or the START field has been selected, a symbolic pressure diagram of the
entire process is displayed on the screen with the current program progress highlighted
against a background and information on the next process steps in the upper left corner of
the screen. During the program run, the screen displays the temperature of the selected
sterilization program or the current temperature in the process chamber in the lower left
corner, the current pressure in the chamber in the lower right corner, and the time remaining
until the end of the process in the upper right corner of the screen. This is an estimated
time, which may be increased due to the weight and type of the load.
During the course of the program, in place of the START field, the STOP field
is displayed, which allows the user to stop the process at any time. In the
upper left corner of the screen, the names of the individual stages of the
program are displayed consecutively, e.g.
chamber lock, heating of the working chamber
Program
stage
Temperature of the
selected process
Approximate time to the
end of the process
The current temperature
in the chamber
Current pressure in
the chamber
Program flow
diagram
STOP
button
If the process was successful, the display alternately shows
information screens about the end of the process and the
sterility of the load and about the possibility of opening the
chamber of the device. By pressing the FINISH field, we go to
the welcome screen.
WARNING! After the process, the chamber, tray and load are hot. Be
especially careful and use protective gloves to remove the load or wait until
it cools down. In the 134C FAST program, instruments are hot and wet after
sterilization.