The course of each performed sterilization is automatically saved on a data carrier (USB stick).
• The memory slot is located on the rear panel of the device.
• It is recommended to periodically archive the data on another medium e.g. computer, laptop.
• The USB flash drive should not be removed from the slot during the process.
USB socket
USB flash drive
7. Data archiving
To ensure proper functioning of the Enbio S / Enbio PRO steriliser, it is recommended that the wearing
parts be replaced according to the schedule below. And periodic inspection of the individual components
of the sterilizer in accordance with the following guidelines.
Name
Replacement frequency
HEPA bacteriological filter
Every 1000 cycles or every 12 months
Connection / drain hose
If damage is observed or once a year
Plugs for water / condensate containers
If damage is observed
Front drawer with seal
Replacement at mandatory service after
1500 cycles/365 days
Enbio Magic Filter
Replace the filter once every 6 months
or sooner when the resin has completely
discolored to an amber color
Element to be controlled
Frequency of inspections
Front gasket
weekly or in the event of incorrect operation
- performed by the user
Bacteriological filter
every week - performed by the user
Connection / drain hose
weekly or in the event of incorrect operation
- performed by the user
Container stoppers
weekly - performed by the user
Enbio Magic Filter
weekly - performed by the user