ELECTRO-METRICS EM-2115 1.0 INTRODUCTION/SPECIFICATIONS
(EM2115S1-5)
TABLE 1.1 CONT'D
EM-2115 SPECIFICATIONS
FUNCTION/CHARACTERISTICS
DESCRIPTION/PERFORMANCE
Amplitude Display:
3-digit ½-inch LED with 1 dB resolution plus
uncalibrated 2-inch analog meter for ease in tun-
ing.
Detector Functions
Peak:
Direct-peak response, calibrated in terms of rms
equivalent sinewave. Associated dump circuitry
controls sample time.
Quasi-Peak (CISPR):
Refer to Table 1.4, Page 1-
Average:
Video averaging,nominal 150 ms time constant.
Crest Factor: 50 dB minimum.
RMS:
True rms of detected video signal, nominal 150
ms time constant.
Crest Factor: 30 dB minimum.
Slideback:
Metered threshold biased envelope detector for
visual or aural extinction of peak levels.
Calibrator:
Impulse generator, 9 kHz-1000 MHz, nominal 80
Hz rate. Automatically activated when Calibra-
tion switch is pushed. Digital amplitude display
automatically compensated.
Spectrum Analysis Controls:
For driving conventional oscilloscope or X-Y
plotter.
Three modes: Manual, Continuous, or single
sweep with variable rate and dispersion.
Sweep times 0.01 to 3000 seconds based on con-
trol settings and band selected, with each setting
variable over a 10:1 range. Dispersion is variable
from 0.1% to 20% (100% at click stop) or the
band selected.
GPIB Interface:
IEEE Std. 488-1978.
Содержание EM-2115
Страница 1: ...INSTRUCTION MANUAL INTERFERENCE ANALYZER MODEL EM 2115 EMC 30...
Страница 68: ...EM 2115 3 0 PERFORMANCE CHECK EM2115S3 13...
Страница 104: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 36...
Страница 106: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 38...
Страница 108: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 40...
Страница 110: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 42...
Страница 112: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 44...