EM-2115 APPENDIX E ACCESSORY DATA
(EM2115E-23)
Width:
22.4 cm (8.8").
Height:
14.8 cm (5.8").
Weight:
Approx. 3.6 kg (8 lbs).
E-9.2 EM-4205 Description
The EM-4205 Broadband Power Amplifier is designed to be used whenever signal ampli-
fication is required within a test system.
The minimum gain of 16 dB, from 10 MHz to 1 GHz, is for a 50
impedance and the
maximum safe input level is 2.2 Vrms. If the 2.2 Vrms input level limit is exceeded, the internal
circuitry of the unit may be damaged.
The EM-4205 can also be used as a post-amplifier in combination with the EM-4206
Broadband Amplifier for special broadband, high dynamic range measurements. The two units
combined provide a third-order spurious free dynamic range (SFDR) of 47 dB (Bandwidth: >990
MHz, Overall Noise Figure: 6.0 dB, Gain: 43 dB). This provides an impulse sensitivity of -31
dB(µV/MHz) for the 990 MHz bandwidth.
E-9.3 EM-4205 Applications
The Electro-Metrics Model EM-4205 Broadband Power Amplifier is general purpose
signal amplifier that can be used in susceptibility tests, shielding effectiveness tests, insertion
loss measurements, plus numerous other applications where signal amplification is required.
E-10.0 Additional Optional Accessory Items
In addition to the optional accessory items described in E-2.0 thru E-9.0, the following
accessory items are also available.
a. EM-1105:
Coaxial Cable, 50
.
Connectors: BNC (male)-TNC (male).
Length: 7.6 m (25').
b. EM-1300:
Rack mounting kit for mounting the EM-2115 in a standard 19-
inch instrument rack.
c. EM-1301:
Antenna Mount.
d. EM-1303:
Antenna Clamp Mount.
e. EM-1305:
Antenna Remote Control Cable.
Содержание EM-2115
Страница 1: ...INSTRUCTION MANUAL INTERFERENCE ANALYZER MODEL EM 2115 EMC 30...
Страница 68: ...EM 2115 3 0 PERFORMANCE CHECK EM2115S3 13...
Страница 104: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 36...
Страница 106: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 38...
Страница 108: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 40...
Страница 110: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 42...
Страница 112: ...EM 2115 4 0 SERVICE MAINTENANCE CALIBRATION EM2115S4 44...