STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 126
SER
IA
L
C
OM
M
U
N
IC
A
T
ION
P
R
OT
O
C
OL
5.4.17 SetOptionFunc
Function: Changes the setting value of the input operation port, warning function, and serial
communication time out.
CAUTION
There is an endurance limit of setting changes (no more than 24 times per day for about ten years).
If this limit is exceeded, it not only may become impossible to change the setting but can also cause
pump failure. Do not make setting changes in excess of this endurance limit.
Transmission frame:
*1
PC->SIM Stx
0
0
1
Sp
=
Parameter 1 to 13
Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx
0
0
1
#
Etx LRC
*1 The HEX code of ASCII character ' = ' is "3D".
Parameter
Item
Data format
Remarks
1
Input operation port setting
8-bits hexadecimal coded
ASCII
Default 01
hex
(I/O REMOTE)
2
TMS Option
Enable/Disable setting
8-bits hexadecimal coded
ASCII
Default FF
hex
(DISABLE)
3
[System reservation]
48-bits hexadecimal
coded ASCII
*1
4
Second Damage Limit Option
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
5
First Damage Limit Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
6
Pump Runtime Over Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default FF
hex
(DISABLE)
7
Pump Runtime Over Warning
Hours setting (×100 hours)
32-bits hexadecimal
coded ASCII
Default
00000000
hex
8
Imbalance Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
9
Pump Overload Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default FF
hex
(DISABLE)
10
Pump Overload Warning
Motor current setting
16-bits hexadecimal
coded ASCII
Default 03E8
hex
(100.0%)
Содержание STP-iXA3306B
Страница 12: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page x PAGE x This page intentionally blank ...
Страница 24: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 12 INTRODUCTION 1 This page intentionally blank ...
Страница 66: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 54 INSTALLATION 3 This page intentionally blank ...
Страница 176: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 164 MAINTENANCE AND INSPECTION 8 This page intentionally blank ...