STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 121
SER
IA
L
C
OM
M
U
N
IC
A
T
ION
P
R
OT
O
C
O
L
5.4.15 ReadMeasValue
Function: Reads the TMS temperature, motor temperature, motor current, measured rotational speed,
and control unit temperature. The motor temperature value is the same temperature as
"ReadMotorTemp". The measured rotational speed value is the same as "ReadMeas" parameter 2
(Measured rotational speed).
Transmission frame:
*1
PC->SIM Stx
0
0
1
?
[
Etx LRC
SIM->PC
Ack
PC->SIM
*1
Ack
SIM->PC Stx
0
0
1
Sp
[
Parameter 1 to 9
Etx LRC
*1
The HEX code of ASCII character ’[’ is "5B".
Parameter
Item
Data format
Remark
1
[System reservation]
120-bits hexadecimal
coded ASCII
2
TMS temperature
(Unit ºC)
16-bits hexadecimal
coded ASCII
3
Motor temperature
(Unit ºC)
16-bits hexadecimal
coded ASCII
4
[System reservation]
8-bits hexadecimal
coded ASCII
5
Motor current
(Unit: 0.1 A)
8-bits hexadecimal
coded ASCII
6
[System reservation]
24-bits hexadecimal
coded ASCII
7
Measured rotational speed
(Unit: Hz)
16-bits hexadecimal
coded ASCII
8
[System reservation]
48-bits hexadecimal
coded ASCII
9
Control unit temperature
(Unit: °C)
16-bits hexadecimal
coded ASCII
Содержание STP-iXA3306B
Страница 12: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page x PAGE x This page intentionally blank ...
Страница 24: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 12 INTRODUCTION 1 This page intentionally blank ...
Страница 66: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 54 INSTALLATION 3 This page intentionally blank ...
Страница 176: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 164 MAINTENANCE AND INSPECTION 8 This page intentionally blank ...