STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 113
SER
IA
L
C
OM
M
U
N
IC
A
T
ION
P
R
OT
O
C
O
L
5.4.8
ReadSetPoint
Function: Reads the setting value of the "Speed Set Point" and TMS temperature. The "Speed Set
Point" data is the same data as that of "ReadSpeedSetPoint".
Transmission frame:
PC->SIM Stx
0
0
1
?
d
Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx
0
0
1
Sp
d
Parameter 1 to 2
Etx LRC
Parameter
Item
Data format
Remarks
1
Speed Set Point
(Unit: Hz)
16-bits hexadecimal
coded ASCII
2
TMS temperature setting
(Unit: ºC)
16-bits hexadecimal
coded ASCII
Example:
Speed Set Point
: 01F4
hex
= 500 Hz = 30,000 rpm
TMS temperature setting
: 0046
hex
= 70 ºC
Parameter
1
2
ASCII
0
1
F
4
0
0
4
6
HEX
30 31 46 34 30 30 34 36
5.4.9
ReadMotorTemp
Function: Reads the measured motor temperature.
Transmission frame:
PC->SIM Stx
0
0
1
?
e
Etx LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC Stx
0
0
1
Sp
e
Parameter 1
Etx LRC
Содержание STP-iXA3306B
Страница 12: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page x PAGE x This page intentionally blank ...
Страница 24: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 12 INTRODUCTION 1 This page intentionally blank ...
Страница 66: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 54 INSTALLATION 3 This page intentionally blank ...
Страница 176: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 164 MAINTENANCE AND INSPECTION 8 This page intentionally blank ...