STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 85
SER
IA
L
C
OM
M
U
N
IC
A
T
ION
P
R
OT
O
C
O
L
5.2.5
Serial communication timeout setting
If the signal to the input operation port of the STP pump is interrupted for a certain period during
acceleration or normal operation, the STP pump detects a failure and stops. The time setting of the
failure detection is user definable. When setting the value to 0, the function is disabled. This value will
be common to all serial ports, and the factory setting is 1 minute.
The setting value can be changed via serial communication, the STP-Link (optional accessory). (see
Design the user application so that the PC can communicate with the STP pump at fixed regular
intervals within the setting time, except when the function is disabled (the value is 0).
Default
Setting Range
Remark
Serial
communication
time out setting
1 minute
0 to 500 minutes
(1 minute step)
The function is disabled when the
value is set to 0.
Table 31 - Serial communication time out setting
Note:
When the communication time out is disabled, the STP pump may not stop when the serial
communication does not function normally due to a breakage of the communication cable. In
this case, interrupt the power supply for 2 seconds or more to stop the STP pump by power
failure detection. Supply the power to the STP pump immediately after power failure
detection.
Содержание STP-iXA3306B
Страница 12: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page x PAGE x This page intentionally blank ...
Страница 24: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 12 INTRODUCTION 1 This page intentionally blank ...
Страница 66: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 54 INSTALLATION 3 This page intentionally blank ...
Страница 176: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 164 MAINTENANCE AND INSPECTION 8 This page intentionally blank ...