STP-iXA3306 Series Turbomolecular Pump
MT-82E-001-H
Page 25
IN
ST
A
L
L
A
T
ION
3.2
Unpacking
3.2.1
Unpacking the STP pump
Check outer package for damage and that the delivery note corresponds to the purchase order. If the
STP pump is damaged, contact Edwards or their distributor.
Note:
It is recommended to keep the packaging materials, such as the corrugated fiberboard
container and cushioning material for possible reuse.
WARNING
The STP pumps are heavy products. Observe national laws/regulations, safety
standards and manufacturers instructions when lifting the STP pump.
Lifting devices must be used when lifting or moving the STP pump.
The STP pumps are heavy products (Refer to Section 2, "TECHNICAL DATA" for the mass of the
STP pump). Use a crane or other appropriate means sufficient enough to withstand the load when
lifting the STP pump of 18 kg or more.
Lift the STP pump with an eyebolt or a similar tool that uses the fitting hole attached to the inlet port
flange. When lifting the pump, use 2 or more ropes. (see Figure 10)
A crane and eyebolts due to lifting operations should withstand the load of five times or more the
weight of the STP pump, and rope should be seven time or more. Use an eyebolt which confirms to
ISO03266.
Always lift the STP pump in stable places free of excessive shock or vibration to prevent it from
shaking or dropping.
Care should be taken not to scratch the flange of the STP pump. Before installing the STP pump,
check that there are no scratches on the surface.
Eyebolt size
Flange type
M8
ICF305, ICF356
M10
ISO250F
M12
VG250, VG300,
VG350, ISO320F
Figure 10 - Example of lifting the STP pump
Table 6 - Eyebolt size
Содержание STP-iXA3306B
Страница 12: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page x PAGE x This page intentionally blank ...
Страница 24: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 12 INTRODUCTION 1 This page intentionally blank ...
Страница 66: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 54 INSTALLATION 3 This page intentionally blank ...
Страница 176: ...STP iXA3306 Series Turbomolecular Pump MT 82E 001 H Page 164 MAINTENANCE AND INSPECTION 8 This page intentionally blank ...