![Edwards nEXT200200 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/edwards/next200200/next200200_instruction-manual_2366136091.webp)
9. Storage
If possible, do not store the pump for long-term. When long-term storage is necessary,
the pump should be set up and operated for minimum eight hours every six months.
To store the pump:
1.
Put protective covers over the inlet, outlet, interstage (for the 'i' nEXT only), purge,
booster (for the 'T' nEXT only) and vent ports.
2.
Put the pump in its packing materials. Seal the pump in a plastic bag with a
applicable desiccant for fastest pump-down when the pump is put back into
service.
3.
Store the pump in cool, dry conditions until necessary for use. Refer to
Operation and storage environment on page 22
for recommended storage
environment.
4.
When necessary, prepare and install the pump as given in
5.
Always keep the pump upright to prevent the drainage of oil from the bearing
reservoir.
Page 87
B85200880_D - Storage
Содержание nEXT200200
Страница 6: ...This page has been intentionally left blank...
Страница 24: ...Figure 3 nEXT 200200 NW25 interstage port pumping speed Page 20 B85200880_D Technical data...
Страница 77: ...6 8 Bakeout nEXT200200 split flow pumps are not applicable for bakeout Page 73 B85200880_D Operation...
Страница 100: ...edwardsvacuum com...