![Edwards nEXT200200 Скачать руководство пользователя страница 86](http://html1.mh-extra.com/html/edwards/next200200/next200200_instruction-manual_2366136086.webp)
8.6 Bearing run time
The run hours and recommended service time of the bearing in the pump can be
monitored.
Send the query that follows:
Command
?
V
8
8
5
cr
The reply is as follows (where, first number is the hours run by the bearing and second is
the number of hours until service is recommended):
Reply
=
V
8
8
5 sp d
d
d
d
d
d
;
d
d
d
d
d
d
cr
8.7 Oil cartridge run time
The run hours and recommended service time of the oil cartridge in the pump can be
monitored.
Send the query that follows:
Command
?
V
8
8
6
cr
The reply is as follows (where, first number is the hours run by the oil cartridge and
second is the number of hours until service is recommended):
Reply
=
V
8
8
6 sp d
d
d
d
d
d
;
d
d
d
d
d
d
cr
8.8 Flashing error codes
When a fail condition becomes active, the red alarm lights continuously or shows a
flashing sequence.
If the error light is on continuously, this shows a problem with the embedded software.
Restart the system. If restarting the system does not clear the fault, a software download
is necessary. Contact us or the supplier. If the alarm LED is flashing, identify the error
flash code and refer to
Table: Flashing error codes on page 82
Sufficient off period is there between each subsequent cycle repetition to mark the start
of a new flash sequence. The duration of a long flash (L) is equal to 3 times the duration
of a short flash (0.5 s).
Table 25
Flashing error codes
Error flash
position
Error flash
code
Comments
Actions
0
ssssss
The speed decreases below
50% of the full rotational
speed with the Timer
disabled.
Check if the pump is too hot
or if the inlet pressure is too
high.
1
Lsssss
Controller internal software
mismatch
Restart the system. If the
error code is not cleared,
contact us or the supplier.
Page 82
B85200880_D - Fault finding
Содержание nEXT200200
Страница 6: ...This page has been intentionally left blank...
Страница 24: ...Figure 3 nEXT 200200 NW25 interstage port pumping speed Page 20 B85200880_D Technical data...
Страница 77: ...6 8 Bakeout nEXT200200 split flow pumps are not applicable for bakeout Page 73 B85200880_D Operation...
Страница 100: ...edwardsvacuum com...