with the help of a screwdriver. The mating plug for the 3
‑
way controller connector is
supplied with the pump.
The connector supplies the power to the vent valve or fan connected to the 3
‑
pins. The
connector mating plug is shown in
Figure: Controller connector showing pin numbers on
, with the polarity of the pins marked, when the vent valve or fan is set to on.
The 24 V d.c. output on the controller connector is protected from the overload or short
circuit to 0 V d.c. / 24 V d.c. Rtn / chassis. The output supplies the current given in
Controller technical data on page 30
. For higher currents, the output voltage will
decrease and can set to off to limit the current.
Table 10
Controller technical data
Description
Data
Connector plug
Phoenix part number FK
‑
MC1881338
Voltage output
24 V d.c. -25%, +10% (18 V d.c. to 26.4 V d.c.)
Current output
500 mA
Figure 8 Controller connector showing pin numbers
Pin number
Signal
Polarity
1
24 V d.c. Out
+
2
Chassis
3
24 V d.c. Rtn
-
3.11 Indicator LED's
The pump has three indicator LED's, shown in
Figure: Controller status information on
.
Table 11
Indicator LED’s
LED
Description
Normal LED
The green LED stays when the rotational speed of the pump is more
than the normal speed setting, irrespective of the acceleration or
deceleration of the pump.
Page 30
B85200880_D - Technical data