Controlled venting options, including operating a fan from the controller
▪
Standby speed setting
▪
Normal speed setting
▪
Timer settings: to enable and disable the timer, setting the timeout period
▪
Electronic braking options
▪
Factory default settings
Refer to the TIC Turbo and Instrument Controller or TIC Turbo Controller Instruction
Manuals for information on how to do the settings.
Note:
You cannot configure the Analogue Output Options using the TIC. You cannot assign a
multi
‑
drop address to the pump with TIC.
Option is available to connect the TIC to a PC and using the TIC PC program to configure
the pump. The TIC PC program allows:
▪
configuration of all settings given in the above list
▪
configuration of the analogue output options
▪
assignment of a multi
‑
drop address.
Refer to the TIC PC program Instruction Manual for more information.