Hexadecimal
Binary
Decimal
1
0001
1
2
0010
2
3
0011
3
4
0100
4
5
0101
5
6
0110
6
7
0111
7
8
1000
8
9
1001
9
A
1010
10
B
1011
11
C
1100
12
D
1101
13
E
1110
14
F
1111
15
Each binary digit (bit) represents a flag that is either active (state 1) or not active
(state 0). To help decode the system status word, each bit is numbered (starting with 0
for the least significant to 31 for the most significant) as shown below.
Refer to
for the list of the lower 16 status flags that are
useful for fault finding. The upper 16 status flags are reserved by us.
Binary digits
0 0 1 0 0 0 1 0 1 0 0 0 0 0 1 1 0 0 0 0 0 0 0 0 0 0 1 0 0 0 1 0
ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê ê
3
1
3
0
2
9
2
8
2
7
2
6
2
5
2
4
2
3
2
2
2
1
2
0
1
9
1
8
1
7
1
6
1
5
1
4
1
3
1
2
1
1
1
0 9 8 7 6 5 4 3 2 1 0
Bit numbers
Table 27
Status flag
Bit number Status flag
Active flags means
0
Fail
Fail status condition active
1
Stopped speed
Below stopped speed
2
Normal speed
Above normal speed
3
Vent valve closed
Vent valve energised
4
Start
Start command active
5
Serial enable
Serial enable active
6
Standby
Standby active
7
Half full speed
Above 50% full rotational speed
8
Parallel control mode
Exclusive control mode selection
9
Serial control mode
Exclusive control mode selection
Page 84
B85200880_D - Fault finding
Содержание nEXT200200