7. Maintenance
WARNING: PERSONAL PROTECTIVE EQUIPMENTS
Risk of injury. When you remove the pump, make sure to use the personal protective
equipment to prevent the damage from failed components and possible
accumulation of hazardous material. Check that the pump is vented to atmosphere at
a safe temperature. Be careful from possible spillages, sharp edges and debris.
Removal of the pump inlet screen will expose the risk of injury from sharp edges.
WARNING: MAINTENANCE SAFETY
Allow the pump rotor to stop before you disconnect the logic interface cable from the
power supply. Isolate other power sources. Remove the pump from the vacuum
system for maintenance or fault finding procedures.
7.1 Introduction
Instructions for bearing maintenance and surface cleaning for the turbomolecular
vacuum pump are given in the sections that follows. The inlet screens, inlet strainer and
inlet flange seals are available as spares. Refer to
.
7.2 Bearing and oil cartridge maintenance
Replace the oil cartridge and oil-lubricated bearing when its service life ends. This is
approximately more than 17,500 hours for the oil cartridge and 35,000 hours for the
bearing.
When the oil cartridge or bearing are necessary to replace, we recommend to:
▪
send the pump to our Service Centre for a bearing or oil cartridge replacement.
▪
purchase an oil cartridge or bearing service kit and replace the bearing or oil
cartridge on-site.
Refer to
.
7.3 Rotor life
The fatigue life of turbomolecular pump rotors is approximately 40,000 to 50,000 cycles.
As a precautionary measure, we recommend you to return the pumps for a major
service (rotor replacement) after 20,000 cycles of acceleration to full speed and back to
a stop, or after ten years of use, whichever occurs first.
Page 74
B85200880_D - Maintenance
Содержание nEXT200200
Страница 6: ...This page has been intentionally left blank...
Страница 24: ...Figure 3 nEXT 200200 NW25 interstage port pumping speed Page 20 B85200880_D Technical data...
Страница 77: ...6 8 Bakeout nEXT200200 split flow pumps are not applicable for bakeout Page 73 B85200880_D Operation...
Страница 100: ...edwardsvacuum com...