To assign a multi
‑
drop address, send the command that follows (where the 'd' characters
represent the address):
Command
!
S
8
5
0
sp
d
d
cr
Note:
The address can be a decimal number from 1 to 98. The address number 0 is used to
disable multi
‑
drop mode. The address number 99 is reserved as a wild card and is used in
the query set up detailed later.
The reply will be as follows:
Reply
*
S
8
5
0
sp
r
cr
The multi
‑
drop address is stored in the pump.
To check if the pump has a multi-drop address, send the command that follows:
Reply
?
S
8
5
0
cr
If the reply is as follows, the pump has multi
‑
drop mode disabled:
Reply
=
S
8
5
0
sp
0
cr
If the pump has a multi
‑
drop address, there will be no reply. Communicate with the
pump in multi
‑
drop message protocol. Refer to
Multi-drop operation on page 50
for
more information about multi
‑
drop mode and multi
‑
drop message protocol.
To check the multi-drop address of the pump, send the query as follows (using wild card
address 99 which means 'any' node):
Command
#
9
9
:
9
9
?
S
8
5
0
cr
The reply will be as follows, where dd shows the multi
‑
drop address of the pump:
Reply
#
9
9
:
9
9
=
S
8
5
0
sp
d
d
cr
Multi
‑
drop mode can be disabled by assigning the pump an address 0. To do this, send
the command that follows (where dd shows the multi
‑
drop address of the pump and xx
shows the address of the node that is sending the command):
Command
#
d
d
:
x
x
!
S
8
5
0
sp
0
cr
The reply will be as follows:
Reply
#
x
x
:
d
d
*
S
8
5
0
sp
0
cr
When the multi
‑
drop mode is disabled, the pump will not respond to multi
‑
drop
commands.
5.2 Configuring the nEXT pump using a TIC
The pump can be configured using the manufacturer's TIC Turbo and Instrument
Controller or TIC Turbo Controller. Refer to
Connect the logic interface to the TIC on page
The parameters of the pump that can be set using the TIC are:
▪
Power limit setting
Page 62
B85200880_D - Configuration