ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
50
3.3.5. Turn on/off auxiliary level
If the classification results depend on auxiliary parameter (secondary parameter), the
user can set the auxiliary parameter limits. Users have two ways for classification based
on the auxiliary parameter.
Assign the auxiliary parameter limits and turn off the classification of the auxiliary
level
Only the DUT with the auxiliary parameter within the auxiliary parameter limits can be
classified based on the comparison result of the main parameters. The DUT with the
auxiliary parameter outside the auxiliary parameter limits will be classified into OUT OF
BIN.
Turn off the auxiliary level
Assign the auxiliary parameter limits and turn on the classification of the auxiliary
level
The DUT with the parameter outside the main parameter limits will be classified into
OUT OF BIN. In addition, when the parameters are within the limits of the main
parameters, but the auxiliary parameter outside the parameter limits, the DUT will be
classified into the auxiliary level.
Turn on the auxiliary level
The steps of turning on/off the auxiliary level:
1. Press the [Meas Setup] key.
2. Press the [Limit Setup] function key.
3. Use the arrow key to select the "Aux", and press the [Enter] key.