ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
28
The following parameters can be set in the correction page
open circuit correction (open)
short circuit correction (short circuit)
load correction (load)
select cable length (cable)
load correction measurement function (function)
frequency point for open, short and load correction (frequency)
the measurement value and reference value at the specified frequency point
for open circuit, short-circuit and load correction (reference A, open A, short circuit A, load
A, reference B, open B, short circuit B, load B)
3.2.1 Open circuit correction
ET35’s open circuit correction function can eliminate the error caused by stray
admittance (G, B) in parallel with the test element. Stray admittance model is as follows
ET35 uses the following two kinds of data for open circuit correction.
First: regardless of what the frequency is set currently, ET35 will conduct the test of