ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
30
Press the [Means Open] function key, and the page is shown below
The steps of open circuit correction based on specified frequency point:
1. Press the [Meas Setup] key.
2. Press [Correction] function key.
3. Use the arrow key to select Spot No to specifiy frequency point.
4. Enter the number of measuring point, and press the [Enter] key to confirm.
5. Use the arrow key to select "Frequency", and press the [Enter] key.
6. Use the numeric key to input frequency, and then press the unit function key to
enter.
7. Connect the test terminal and the test fixture without connecting to any DUT.
8. Press the [Open Correction] function key.
9. The correction measurement results are displayed after "open A" and "Open B".
10. Use the arrow key to select "Open", and press the [Enter] key.
11. Press the [On] function key to initiate the calculation of open circuit correction at
Содержание ET3501
Страница 25: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 22...
Страница 77: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 74...
Страница 93: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 90...