BIN Count display
Counter function
BIN can be counted by ET35. ET35 can classify the components into the appropriate
BIN with the counter function when counting the components classified into each BIN.
999999. is the limit. When the count reaches the maximum value, counting stops and
the message of overflow "----" is shown.
Steps of setting the counter:
1. Press the [Display Format] key.
2. Press the [BIN Count] function key.
3. Use the arrow key to select the "count".
4. Use the function key to turn the comparator on or off.
Steps of clearing the counter:
1. Press the [Display Format] key.
2. Press [BIN Count] function key.
3. Use the arrow key to select the "Count".
4. Press the [Clear] function key.
(4) List sweep Display
When you press the [Display Format] key and [List sweep] function key, the List
sweep page will be displayed. The user can observe up to 10 points of test frequencies,
and measure the level or DC bias. The upper and lower limits can be set for each test
point for the List sweep. These test points are automatically swept and tested; the test
results will be compared with their corresponding limits. The limits cannot be set on this
page but on the List Settings display.
Users can set the sweep mode on this page