ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
35
The steps of short circuit correction based on all frequency points:
1. Press the [Meas Setup] key.
2. Press the [correction] function key.
3. Use the arrow key to select Spot No to specify the frequency.
4. Enter the number of measuring point, and press the [Enter] key to confirm.
5. Use the arrow key to select "Frequency", and press the [Enter] key.
6. Use the numeric key to input frequency, and then press the unit function key to
enter.
7. Short connect the test terminals and test fixture.
8. Press the [short circuit correction] function key.
9. The short circuit correction measurement results are displayed after "Short A" and
"Short B".
10. Use the arrow key to select "short", and press the [Enter] key.
11. Press the [On] function key to initiate the short circuit correction calculation
function.