ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
79
Users can save up to 81,920 groups of measurement results at one time.
Before saving the measurement results to a USB memory, the user needs to do some
preparation work to start the measurement, so that the data buffer contains the required
data.
Format of measurement results
Data A and Data B in the measurement result file vary depending on whether the
“valid digit” setting is on.
Comp
arator
Format of result
Off
<Data A>, <Data B>, <Status>
On
<Data A>, <Data B>, <Status>, <BIN No.>
<Data A> and <Data B> adopt the scientific notation with the precision of
six digits after the decimal point
Status uses one of the following status values to represent the state of the
measurement results:
0 Measurement completed successfully.
+1 Overload.
+3 Signal beyond the permitted limit of the signal source detected.
+4 Automatic level control functions fail.
0
OUT_OF_BINS
+1
~
+9 BIN 1
~
BIN 9
+10
AUX_BIN
BIN No
(
IN/OUT
)
respresents the bin classification result in the measurement as
shown below:
0
OUT_OF_BINS
+1
~
+9 BIN 1
~
BIN 9
+10
AUX_BIN
The steps of saving the measurement results on the USB memory:
1. Insert the USB memory into the front USB port.
2. Press the [Save] key.
3. Press the [Save Data] function key.
4. Press the [Start recording] function key to enter the result into the data buffer
memory.
5. Start measurement.
6. Press the [Save and stop] function key to stop the recording and save the results to
a USB memory.
7. Prompt success.