ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
10
BIN Number Display
Comparator function ON / OFF
The internal comparator of ET35 can use up to nine pairs of primary parameter
limit and a pair of auxiliary parameter limit, and classify the device into up to 10 kinds
(binBIN1 ~ BIN9 and OUT OF BIN). In addition, the device with its first parameter
within the limit range and its auxiliary parameter measurement outside the limit range
can be classified as an auxiliary BIN (AUX BIN).
To use the comparator function to control device process, HANDLER interface is
required.
The above mentioned setting of limits for BIN classification can be done with the
[Meas Setup] on the page of limit list. Therefore, users to compare the filed only need
set the comparator function as on or off.
The step of setting the comparator function:
1. Press the [Display Format] key.
2. Press the [BIN Number Display] function key.
3. Use the arrow key to select "comp".
4. Use the function key to turn the comparator on or off.
(3) BIN Count page
Press the [Display Format] key and [BIN Count] function key, and the page of
BIN Count will be shown. The counting result of the comparator will be shown on the
page.
count function on / off
The following information is also shown in the monitoring area on the page.
These conditions can be set on the page of limit list.
function
nominal value