Although the contact mark may remain on the rotor when
you use the rotor cover (Polypropylene) repeatedly,
there is no problem.
Contact part with the
cover-locking knob
Mounting the rotor cover (AL) on FAS15E/15F/15G
rotor
(1) Apply silicone grease (vacuum grease) (Part
No.483719) to the rotor packing (O-ring)and then
fit the rotor packing in the groove of the cover.
Replace the rotor packing with new one if
damaged or deteriorated.
(2) Apply lubricant for screw (Part No. 84810601) to
the threaded portion of the cove-locking knob.
Place the rotor cover on the rotor body.
Cover-locking knob
O-ring
Rotor cover
Tighten the cover-locking knob to fix the rotor
cover to the rotor body securely.
O-ring
Threaded portion
7. Run the rotor according to the procedures specified in the centrifuge instruction manual.
8. After the operation, gently remove the rotor cover from the rotor. Take out the tubes from the rotor and then
take out the sample in a proper manner.
Precautions in Cleaning
・
Wipe the inside of the drive hole (crown hole) of the rotor with a soft cloth dampened with water
sufficiently.
・
Use tap water or neutral detergent (pH5
~
pH9) for cleaning the rotor and the adapters.
・
If sample is spilled in the rotor, wash the stains off and dry it well to avoid corrosion.
・
In the event a tube is broken while the rotor is running and the fragments or foreign matter remains in the
tube cavities, remove the fragments or foreign matter completely.
Precautions in Sterilization
・
Sterilize the rotor in any of the following methods; autoclaving (121
℃
, 20 minutes), gas sterilization
(ethylene oxide or formaldehyde), chemicals sterilization (ethanol (70%), hydrogen peroxide (3%) or
formalin (3%)), or ultraviolet rays sterilization.
・
Sterilize the adapters in the following method; chemicals sterilization (hydrogen peroxide (3%)).
Sterilizing the rotor cover (material: Polypropylene) of theFAS15E rotor
・
Sterilize the rotor cover (material: Polypropylene) in the following method; chemicals sterilization
(ethanol (70%) and hydrogen peroxide (3%)).
!
CAUTION
:
Be sure to tighten the cover-locking knob securely before operation when mounting the rotor
cover to the rotor body. Otherwise the rotor cover might be removed during operation and
result in damage to the centrifuge and the rotor.
!
CAUTION
:
Do not dip the rotor in the formalin (3%) solution more than 2 hours.
!
CAUTION
:
Be sure to sterilize the rotor after removing the rotor cover from the rotor.
!
CAUTION
:
Inspect the cover-packing (O-rings) after sterilization. If you observe deterioration of
the cover-packing (O-rings), replace them with new ones.
!
CAUTION
:
Never sterilize the rotor cover by autoclaving. Otherwise the rotor cover might be
deteriorated.