![Dynamica Velocity 15HR Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262034.webp)
6-2
6. Troubleshooting
(1) When an alarm code is displayed
■
Alarms E-1 to 9 are mainly related to handling errors. You can continue using
the centrifuge if the cause is removed.
If alarm codes can not be cleared by
taking the proper action, contact a Dynamica authorized sales/service
representative.
■
Alarm E-13 indicates a rotation sensor error. Once this alarm code is displayed,
the alarm code cannot be cleared and the door cannot be opened for six minutes
– a time that allows all the rotors to come to a halt with the coasting deceleration.
Wait at least six minutes keeping the power on and press the DISP/CE button
to remove the rotor.(When turning off the POWER switch, you can not clear the
alarm code for six minutes after turning on the POWER switch.)
■
Alarm E-15 indicates a motor over-temperature. When the room temperature is
high or the rear of the centrifuge is covered, this alarm code is displayed even if
the centrifuge is not out of order. Lower the room temperature or keep
clearances of at least 30 cm behind the centrifuge. And then clear the alarm
code.
■
Alarm E20 and E21 are indicated when the drive shaft is rotated at high speed
without the rotor. When these alarm codes are indicated, turn off the POWER
switch to clear the alarm code.
(2) How to clear alarm codes
■
Alarm codes can be cleared by pressing the DISP/CE button after removing
the cause.
■
When the alarm code is displayed during operation, the door is not unlocked
automatically even though the rotor has stopped. After clearing the alarm code,
unlock the door by pressing the OPEN button.
■
If alarm codes can not be cleared by taking the proper action, contact a
Dynamica authorized sales/service representative.