試料の平均密度(g/ml)
1.2
・
This rotor can separate sample whose mean density is 1.2
g
/mL or less.
When the mean density of the sample to be separated is over 1.2
g
/
mL, calculate the allowable speed
according to the following equation.
Allowable speed (rpm)
=
15,000(rpm)
×
・
To find the RCF: Substitute the rotation radius (Rmax., Rmin., etc.) and the speed in the following equation.
RCF (x g) = 1.118 x 10
-5
x Rotation radius (cm) x (Speed (rpm))
2
Applicable Tubes and Adapters
Maximum speeds listed are guidelines only. Because of variances in user methodologies, no guarantee of
performance is expressed or implied.
FAS15E
Adapter
Tube
Max. speed
(Max. RCF)
Part No.
Name
Nominal
capacity
Part No.
Name
Actual
capacity
Size (mm)
2.0 mL
(Marketed)
*
2.0 mL micro tube
1.7 mL
(D) 10.8
×
(L) 39
15,000 rpm
(21,500 xg
*
3
)
1.5 mL
(Marketed)
*
1.5 mL micro tube
1.4 mL
(D) 10.8
×
(L) 39
S413789A
1GA2 adapter
(2 pcs.)
(Material:
*
1)
1.0 mL
489241A
1.0 round-bottom glass tube
(100 pcs.)
0.7 mL
(D) 9.0
×
(L) 35
0.7 mL
(Marketed)
*
0.7 mL micro tube
0.6 mL (D) 10.8
×
(L) 25
S413789A
1GA2 adapter
(2 pcs.)
(Material:
*
1)
0.5 mL
489240A
0.5 glass tube (100 pcs.)
0.5 mL
(D) 9.0
×
(L) 35
487293A
0.75 adapter
(4 pcs.)
(Material:
*
2)
0.5 mL
(Marketed)
*
0.5 mL micro tube
0.5 mL
(D) 7.8
×
(L) 30
487292A
0.4 adapter
(4 pcs.)
(Material:
*
2)
0.4 mL
(Marketed)
*
0.4 mL micro tube
0.4 mL
(D) 5.7
×
(L) 46
487292A
0.4 adapter
(4 pcs.)
(Material:
*
2)
0.25 mL
(Marketed)
*
0.25 mL micro tube
0.25 mL
(D) 5.7
×
(L) 30
(Marketed)
*
Centrifugal filter unit
(D) 12.3
×
(L) 45
Unit (mm
)
L
L
D
*
: When using tubes on the market, perform operation under the allowable speed or the allowable RCF specified by
the manufacturer. Otherwise the tubes may be broken during operation.
*
1: PP
*
2: ABS
*
3: Max. RCF is 21,500xg, when Rmax. is 8.56cm.
!
CAUTION
:
Use the FAS15E/FAS15F/FAS15G rotor only in the Velocity 15HRE centrifuge.
These rotors can not be used in other centrifuges.
!
CAUTION
:
When using tubes outside 4-25
℃
, tubes may be broken or deformed during operation.
Test them before actual operation and perform actual operation after checking that
there is no inconvenience such as the above.
1.2
(g/mL)
Mean density of sample (
g
/mL)