2-1
2. Installation
This section describes instructions that you need to follow when installing the centrifuge.
The instructions ensure your safety and maximum performance.
(1) Location
Position this centrifuge on a firm and level floor. Avoid installing it on a slippery
floor, or on a floor that conveys vibrations (for example, from people’s footsteps).
This centrifuge can be used at the ambient temperature range of 2°C to 32°C.
We recommend that this centrifuge will be used at the ambient temperature
range of 15°C to 25°C to ensure maximum performance.
Do not position this centrifuge in direct sunlight.
This centrifuge has an air-cooling system. To secure efficient cooling, keep clearances
of at least 10 cm on both sides of the centrifuge, and at least 30 cm behind it.
Do not install near a heat-generating device or water source from where water
might drip or splash. Such a location might increase the temperature of samples
or cause the centrifuge to malfunction.
Do not dew the centrifuge. When moving the centrifuge from the cold room to the
ordinary room, perform a) or b).
a) Warm up the centrifuge by performing the continuous operation for more than 30 minutes.
b) Place the centrifuge for more than three hours in a dried room without
connecting the power cord into the inlet of the centrifuge.
CAUTION
:
●
Before moving the centrifuge, the rotor must be removed.
!
On a firm and level floor.
A room with a good ventilation
A room that is not exposed to the direct sunlight
Room temperature:
From 15
℃
to 25
℃
Humidity
From 20
%
to 90
%
A room with no chemicals and no inflammable gas
A place where water do not drip or splash
Noncondensing
(Be careful when moving the centrifuge
from a cold room to a ordinary room.)
25
℃
15
℃
90
%
20
%