4-6
4. Operation
4.2 Normal Operation
(1) Turn on the POWER switch.
■
All indicators on the operation panel stay lit for about 3 seconds for self-checking.
(2) Open the door and then set tubes into the rotor (see Chapter 3).
■
Door lock is released by pressing the OPEN button while the door is locked.
(Even if pressing the OPEN button, the door lock can not be released when the
rotor is rotating or the door is open.)
(3) Close the door to start running.
■
When closing the door of the centrifuge like pressing the door lightly, the door is
locked automatically. If you need to open the door again, press the OPEN button
to unlock the door.
NOTE
:
●
As soon as the power switch is turned on, the function of self-checking works.
Any button does not work during self-checking.
Velocity 15HR (the right side of the centrifuge)
CAUTION
:
●
Do not insert your fingers or a object into the door hook hole. If the door lock
functions by inserting your fingers into the door hook hole, you might injure your
fingers with the door lock of this centrifuge. Malfunction of this centrifuge might be
caused by inserting a object into the door hook hole.
●
Be careful not to get your fingers or hands caught between the door hook and
the table when closing the door.
!