6-3
6. Troubleshooting
6.2 Troubleshooting when no alarm code is displayed
If the centrifuge cannot operate properly even though no alarm code is displayed,
take the corrective actions specified in the table below.
Symptom
Probable cause
Action
Nothing appears on
the display when the
POWER is turned
on.
Building power circuit
breaker trips or the fuse is
blown out.
Remove the cause of the trouble and turn on
the circuit breaker or replace the fuse.
If alarm codes can not be cleared by taking
the proper action, contact a Dynamica
authorized sales/service representative.
Building power
circuit breaker trips
or the fuse is blown
out while it is
running.
Power is wired from a single
outlet to multiple devices.
The power capacity may be exceeded.
Supply power for this centrifuge and other
devices from different sources.
The centrifuge may be out of
order.
Set the rotor and start operation to check for
successful rotation (look at the SPEED/RCF
display). If normal, continue using the
centrifuge.
CAUTION
:
Turn off the power if any
abnormality is found on the
centrifuge and contact
a Dynamica authorized
sales/service representative.
The centrifuge
does not accept
the entries on
the operation
panel.
The centrifuge may be out of
order.
When the rotor does not stop even if pressing
the STOP button, turn off the POWER
switch and turn on the switch again. Take out
the rotor after the rotor stops.
Contact a Dynamica authorized
sales/service representative.
!