5. Load the balanced tubes and adapters in the rotor symmetrically with respect to the center of rotation of the
rotor .
!
CAUTION
:
Do not load the tubes asymmetrically. Asymmetrical loading may cause imbalance
operation and the centrifuge may be damaged.
Examples of incorrect tube setting are shown below.
When using the FAS15E/FAS15F rotor
Center of rotation
●
:
Tubes are loaded.
○
:
Tubes are not loaded.
When using the FAS15G rotor
Center of rotation
6. Mount the rotor cover to the rotor. This rotor is available without setting the rotor cover to the rotor. It is
recommended that the rotor cover is set to the rotor in order to prevent contamination of the sample with
dust or water and avoid dispersing the sample to the rotor chamber when the sample leakage occurs.
We recommend that you mount the rotor cover to the rotor because it will prevent the temperature of the
sample from increasing and it will prevent the noise of the centrifuge from increasing during the operation.
Mounting the rotor cover (Polypropylene) on FAS15E rotor
(1) Apply silicone grease (vacuum grease) (Part No.483719)
to the rotor packing (O-ring) and then fit the rotor packing
in the groove of the cover-locking knob.
Rotor cover
Cover-locking knob
O-ring
Replace the rotor packing with new one if damaged or
deteriorated.
(2) Apply lubricant for screw (Part No. 84810601) to the inner
claw of the cover-locking knob.
(3) Set the rotor cover onto the rotor. Adjust UNLOCK of the
cover-locking knob to the match marks of the rotor. Then,
push down the cover-locking knob and rotate it for adjusting
LOCK to the match marks of the rotor.
(4) Check that the rotor cover cannot be rotated.
Match marks (Rotor)
Inner Claw
Rotor cover
Cover-locking knob
①
②
①
②
Rotor
!
CAUTION
:
In using FAS15E rotor whose cover is made of polypropylene, hold the rotor body when
tightening the cover-locking knob. Holding the rotor body avoids moving the rotor body.
!
CAUTION
:
Be sure to adjust LOCK (of the cover-locking knob) to the match marks of the rotor.
Otherwise the rotor cover might be removed during operation and result in damage
to the centrifuge and the rotor. If you can rotate the rotor cover after adjusting LOCK to
the match marks of the rotor, the rotor cover may have deteriorated or been
transformed. In that case, replace it with new one.