![Dynamica Velocity 15HR Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262033.webp)
6-1
6. Troubleshooting
6.1 Corrective actions when alarm code is displayed
An alarm code is indicated on the SPEED/RCF display of the operation panel with a
beep if any abnormality is found in the centrifuge. Take the corrective actions
specified in the table below.
Alarm code
Probable cause
Action
E 1
POWER
・
While the rotor was rotating, a
power failure occurred or power
supply voltage was decreased.
・
The centrifuge stops the rotor upon
recovery from the power failure.
Retry operation as needed.
・
Contact a power company if
abnormality is found about the
power supply voltage.
E 2
DOOR
・
The START button is
pressed while the door is open.
・
Press the START button after
closing the door.
・
The door is opened while the rotor
is rotating. ( E-48 might be
indicated.)
WARNING:
:
Never unlock the door while the
rotor is rotating.
E 4
TEMP
・
Cooling ability is degraded.
・
Lower the room temperature. If there
is a heat source nearby, move the
centrifuge away from it.
・
The air inlet of the centrifuge is
blocked
・
Keep clearances of at least 10 cm
from the air inlet of the centrifuge.
・
The cap at the tip of the drain
hose removed.
・
Put the cap on.
E 9
IMBALANCE
・
The allowable imbalance is
exceeded.
・
Balance the sample keeping the
allowable imbalance.
・
The centrifuge is shocked during
operation.
・
An alarm code may be indicated when
the centrifuge is shocked during
operation.
E 10
E 99
・
Alarm codes to report control
errors.
・
For the E13 , E15 , E20 , E21 ,
refer to (1).
・
Contact a Dynamica authorized
sales/service representative.
Remember to provide the alarm code
and the situation.
!
WARNING
:
●
Unspecified repairs, remodeling or disassembly of the centrifuge that is not listed
below is strictly prohibited other than a Dynamica authorized service
representative.
●
Close the door immediately if the rotor is still rotating when the door is opened after
releasing the alarm status.
!