![Dynamica Velocity 15HR Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262051.webp)
FAS15G
Single or 8-strip tubes and caps can be used in this rotor.
Adapter
Tube
Max. speed
(Max. RCF)
Part No.
Name
Nominal
capacity
Part No
.
Name
Actual
capacity
Size (mm)
0.2 mL
(Marketed)
*
0.2 mL PCR tube (single)
*
1
(D) 6
×
(L) 24
15,000 rpm
(21,500 xg
*
2
)
0.2 mL
(Marketed)
*
0.2 mL PCR 8-tube strip
*
1
(D) 6
×
(L) 24
(20,100 xg
*
3
)
L
D
Unit (mm
)
*
: When using tubes on the market, perform operation under the allowable speed or the allowable RCF specified by the
manufacturer. Otherwise the tubes may be broken during operation.
*
1: Actual capacities of PCR tubes on the market may vary depending on the specifications. Refer to the manufacture’s
catalog.
*
2: Max. RCF is 21,500xg, when Rmax. is 8.56cm (Corner tube hole).
*
3: Max. RCF is 20,100xg, when Rmax. is 7.98cm (Center tube hole).