![Dynamica Velocity 15HR Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/dynamica/velocity-15hr/velocity-15hr_instruction-manual_2548262043.webp)
8-1
8. Specifications
Velocity 15HR
Maximum speed
15,000 rpm
Maximum RCF
21,500
×
g
Maximum capacity
2 mL x 24 tubes + 0.5 mL x 24 tubes (FAS15E rotor)
Range of speed control
300 to 15,000 rpm in 100-rpm increments
Temperature
-20 to 40
℃
( “Lo” is displayed as -20
℃
)
Refrigerator
Hermetically sealed
(refrigerant: HFC134a)
(For further information, see the instrument labels*
1
.)
Timer
1-99 minutes
(
in 1-minute intervals
)
with HOLD function
Acceleration/deceleration
control
2-stage variable acceleration control, 2-stage braked deceleration,
plus coasting deceleration control
Driving motor
Brushless DC motor
(
inverter-controlled
)
Memory-based
programmed operation
Selectable three programmed operations (SPEED/RCF, TIME, TEMP
(ACCEL, DECEL, and rotor stop signal)
Rotor stop signal
Selectable from seven kinds of sounds including five tunes, electronic
beep, and no sound
Safety devices
Door interlock, dual-overspeed detector, imbalance detector and
detector for over-temperature of motor
Applicable standard
CE marking
Heat dissipation
500 W
Power requirements
120 V instrument:
108-126 V AC, 15 A, 60 Hz
220-230 V instrument:
198-253 V AC, 10 A, 50Hz ;
198-242 V AC, 10 A, 60Hz
Dimensions
290 (W)
×
538 (D)
×
307 (H) mm
Weight
40 kg
Ambient temperature
Ambient Temperature Range
:
2
℃
- 32
℃
Ambient Temperature Range for performance guarantee
:
15
℃
- 25
℃
*1: Because the instrument labels may differ from one country to another, please contact your local
dealer or a Dynamica authorized sales/service representative.
The Velocity 15HR centrifuges satisfy CE marking requirements. The CE marking is an
international symbol, which shows that the product conforms to EU directives.
Standards concerning these directives are as follows:
• Product Safety (EN61010-1 and EN61010-2-020)
Environment requirements:
• indoor use;
• altitude up to 2000 m;
•
maximum relative humidity of 80 % for temperature up to 31
℃
decreasing linearly to 50 %
relative humidity at 40
℃
;
• Electromagnetic compatibility (EN61326-1)
Pollution degree
2
Installation category
Ⅱ
Содержание Velocity 15HR
Страница 2: ......
Страница 58: ...MEMO ...