4.4 Troubleshooting
For any abnormality, rectify the problem by referring to the following table.
No.
Symptom Cause
Solution
1 Lamp fails to light if power
switch is turned on.
a. Fuse is not in place.
b. Fuse is blown.
a. Reinsert the fuse.
b. Replace the fuse.
2 “CALIBRATE
INSTRUMENT”,
appears at the bottom of the
screen during initialization.
Wavelength calibration values
have been deleted.
a. Execute
CALIBRATION.
b. Contact the nearest Dynamica agent if
the error display does not go off even
after executing CALIBRATION.
3 D
2
LAMP or WI LAMP Error is
displayed on the screen
Lamp filament has failed
Replace the lamp.
4 “PRINTER
HARDWARE
ERROR” appears on the LCD
screen.
Printer head is abnormal.
a. Turn off the power switch, and see if the
error display goes off.
b. Contact the nearest Dynamica agent if
the error message is displayed again.
5
Error message appears on the
bottom of the LCD.
Instrument has been operated
erroneously.
a. Operate the instrument properly referring
to the message displayed.
6 Too large variations in
displayed data
a. Lamp
deteriorated.
b. Lamp contaminated with
fingerprints or dust.
c.
Cell or window
contaminated with water
drops or dust.
a. Replace lamp.
b. Remove contamination or water drops.