3-22
3.4.3 Definition of the options
<START WL> and <STOP WL>
The minimum scan range is 10nm. The Action bar-user prompt will guide you
to enter the appropriate value within the range.
Enter the <START WL> or <STOP WL> wavelength by pressing the numeric
key on the keyboard and then press Enter
Press 0 on the keyboard to save the setup. The HALO RB-10 will check the
setup automatically. If the scan range is under 10nm, the HALO RB-10 will not
start until a correct range has been entered.
The HALO RB-10’s default setting is that the <START WL> is larger than
<STOP WL>. The HALO RB-10 will swap the <START WL> and the <STOP
WL> wavelength automatically if the <STOP WL> is larger than the <START
WL>.
Data Mode
Press 3 on the keyboard, the data mode selection will be highlighted. The
Action Bar/user prompt display the section of the data mode.
- <1- %T - % Transmittance>
- <2- ABS – Absorbance>
- <3- E (S) – E >
Press the numeric key to select the data mode
nm
-0.001
A
Wavelength Scan
1
Start WL(nm)
900
2
Stop WL(nm)
800
3
Data Mode
ABS
4
Upper Limit
3.000
5
Lower Limit
-3.000
6
Speed (nm/min)
400
0
End Setting
Select Function:_
500.0
1
%T
2
ABS
3
E(S)